For learning how a pacemaker and a cardioverter-defibrillator work through the ECG and the marker channel: normal operation, malfunctions and the therapy sequence.
Change the mode, rate, AV delay, pulse and sensitivity: a 10-second strip and the marker channel show how the device saw each beat.
Preset scenarios
Mode
Atrial lead
P wave at the lead: 2 mV. The higher the number, the less the device senses.
Ventricular lead
R wave at the lead: 7 mV. The higher the number, the less the device senses.
Ventricular rate
78/min
Longest pause
770ms
Ventricular pacing
100 %
Atrial pacing
0 %
Set the zones and choose an arrhythmia: the device counts fast intervals, delivers ATP first in the VT zone and, if needed, charges and delivers a shock.
Arrhythmia
ATP result
Terminated by ATP
In the model: 18 fast intervals in the VF zone, redetection after ATP 12 intervals, charge 6 seconds. Real values depend on the device and its programming.
The model is simplified: blanking periods, mode switching (mode switch), rate response and SVT discriminators are not shown. The values are a teaching example; real programming is chosen according to the device, the manufacturer and the patient.